Bartec Measuring transducer for Pt100

SKU: 07-7311-93T5 7350

Bartec Measuring transducer for Pt100: The MODEX series mounts a temperature measuring transducer on-site like a rail-mounted terminal. The module transforms the signal from the Pt100 temperature sensor into a proportional, load-independent 4 to 20 mA output signal. The sensor circuit meets Ex protection type Ex ia for intrinsic safety. An output current exceeding the 4 to 20 mA range indicates a sensor fault (open/short circuit). You can operate the Pt100 temperature sensor in 2-wire or 3-wire circuits in Zone 0 or Zone 1.

$93000

Shipping Time: 4-6 weeks

Need faster options? Contact our Hazardous Areas Experts now for quick alternatives!

Trusted Clients

  • Logo PH66
  • Logo  Exxon
  • Logo Bp
  • Logo Conoco Phillipis
  • Logo Irwing
  • Logo Bayer
  • Logo Tesoro
  • Logo Northen TIer
  • Logo Udelhoven
  • Logo Aloha petroleum
  • Logo Rand
  • Logo Arrow Energy
  • Logo Schlumber
  • Logo Afton
  • Logo Holly frontier
Similar Category: Tag:

Bartec Measuring transducer for Pt100

Bartec Measuring transducer for Pt100: The MODEX series mounts a temperature measuring transducer on-site like a rail-mounted terminal. The module transforms the signal from the Pt100 temperature sensor into a proportional, load-independent 4 to 20 mA output signal. The sensor circuit meets Ex protection type Ex ia for intrinsic safety. An output current exceeding the 4 to 20 mA range indicates a sensor fault (open/short circuit). You can operate the Pt100 temperature sensor in 2-wire or 3-wire circuits in Zone 0 or Zone 1.

Specification:

  • Enclosure Material: High-quality thermoplastic
  • Protection Class:
    • Module: IP 66 EN/IEC 60529
    • Terminals: IP 20 EN/IEC 60529
  • Terminals:
    • 2.5 mm², fine stranded
  • Mounting Rail: TH 35 x 7.5 (15) EN/IEC 60715
  • Terminal Designation: Written marking labels
  • Ambient Temperature: -40 °C to +60 °C at T4
  • Storage Temperature: -40 °C to +60 °C
  • Weight: 0.250 kg

Electrical Data

  • Operating Voltage: DC 24 V -10 %, +20 %
  • Power Consumption: 1.6 W
  • Sensor: Pt100 temperature sensor (2- or 3-wire circuits)
  • Output:
    • Load independent current: 4 to 20 mA
    • Max. load: ≤ 400 Ω
  • Temperature Range:
    • -50 °C to +100 °C
    • 0 °C to +200 °C
    • 0 °C to +400 °C
  • Accuracy: ± 1 % of upper value
  • Function Test:
    • Connect 100 Ω resistance to terminal 15 – 16
    • Bridge terminals 16 and 17
    • Apply current between L- and terminal 31

Explosion Protection

Marking:

  • ATEX II 2G Ex db e IIC Gb
  • II M2 Ex db e I Mb
  • IECEx Ex db e [ia Ga] IIC/IIB Gb
  • Ex db e [ia Ma] I Mb
  • CSA Class I, Zone 1, IIC
  • A/Ex d e [ia] IIC Gb

Certification:

  • PTB 97 ATEX 1068 U
  • TÜV 97 ATEX 1204 X
  • IECEx PTB 11.0083U
  • IECEx TUN 11.0030X
  • CSA 2011-2484303U

Certificate:

Bartec Measuring transducer for Pt100 atex certificate

Bartec Measuring transducer for Pt100 IECEX certificate

Bartec Measuring transducer for Pt100 CSA certificate

Datasheet:

Bartec Measuring transducer for Pt100 datasheet