An intrinsically safe temperature data logger, the TEMP1000EX from MadgeTech has been approved for usage in ATEX Zones 0, 1, and 2. It is backed by certifications from IECEx and FM. As far as the temperature range goes, it monitors and records readings from zero degrees Celsius to one hundred and six degrees Fahrenheit (-40 degrees C to plus one hundred and six degrees Celsius). Additionally, it has a stainless steel shell, a Resistance Temperature Detector (RTD) sensor, 32,767 times and date stamped readings, and a flush bottom or key-ring end cap.
| ATEX Rating | Zone 0 |
| IECEx Certified | Yes |
In addition to being intrinsically safe, the TEMP1000EX temperature data recorder from Madge Tech is certified for usage in ATEX Zone 0, 1, and 2 hazardous areas. It is backed by certifications from IECEx and FM. In addition to the stainless steel casing and wide operating temperature range of -40°C to +80°C, these intrinsically safe data loggers have numerous other advantages.
The Temp1000Ex can record up to 32,767 time and date stamped readings and is available in two styles: flush bottom and key ring end cap. It is also more widely accepted, which benefits those who must adhere to rigorous guidelines while working in hazardous environments. These loggers have been designed specifically to ensure the safety of things in explosive environments.
The Temp1000Ex is compatible with the latest MadgeTech 4 Software making starting, stopping, and downloading simple and easy. Graphical, tabular, and summary data is provided for analysis and can be viewed as customizable engineering units.
| General | |
| Memory | 32,767 readings |
| Reading Rate | 1 reading every 2 seconds to 1 reading every 12 hours |
| Battery Life | 2 year typical at 15 minute reading rate |
| Material | 316 Stainless Steel/Radel |
| Dimensions | 2.65 in x .97 in dia. (67.3 mm x 24.6 mm dia.) |
| IP Rating | (Not Rated) Caution: Do not submerge this product to retain IS rating. |
| Operating Environment | -40 °C to +80 °C (-40 °F to +176 °F), 0 %RH to 95 %RH (case properly sealed) |
| Required Interface Package | IFC400 or IFC406 |
| Approvals | CE |
| ATEX | |
| IECEx | |
| FM 3610 | |
| Temperature | |
| Temperature Sensor | Resistance Temperature Detector (RTD) |
| Temperature Range | -40 °C to +80 °C (-40 °F to +176 °F) |
| Temperature Resolution | 0.01 °C (0.018 °F) |
| Temperature Calibrated Accuracy | ±0.5 °C (0 °C to +50 °C), ±0.9 °F (32 °F to 122 °F) |